开放获取期刊获得更多读者和引用
700 种期刊 和 15,000,000 名读者 每份期刊 获得 25,000 多名读者
Shuai Shi, Zhi Chen, Joseph M Rizzo, Andrew Semple, Sarita Mittal, Jason K Cheung, Daisy Richardson, Valentyn Antochshuk and Mohammed Shameem
A highly sensitive method has been developed for the quantitation of polysorbate 20 (PS20) and 80 (PS80) in therapeutic peptide formulations. A mixed-mode HPLC column was used to separate polysorbates from the peptide and other excipients, and a charged aerosol detector (CAD) was used for the detection. The method was capable of reporting polysorbates as low as 5 ppm, and the sensitivity could be further improved on a needed basis. The method has been used to study the compatibility between polysorbates and m-cresol in the peptide formulation. It was found that both PS20 and PS80 are compatible with m-cresol (at 2.8 mg/ml) when their levels were not greater than 20 ppm. Significant losses of polysorbates were observed when PS20 and PS80 concentrations were above 50 ppm. Furthermore, the agitation study demonstrated that even trace levels of PS20 and PS80 (e.g., 20 ppm) could stabilize the peptide against fibrillation and aggregation.