国际标准期刊号: 2155-9872

分析与生物分析技术杂志

开放获取

我们集团组织了 3000 多个全球系列会议 每年在美国、欧洲和美国举办的活动亚洲得到 1000 多个科学协会的支持 并出版了 700+ 开放获取期刊包含超过50000名知名人士、知名科学家担任编委会成员。

开放获取期刊获得更多读者和引用
700 种期刊 15,000,000 名读者 每份期刊 获得 25,000 多名读者

索引于
  • CAS 来源索引 (CASSI)
  • 哥白尼索引
  • 谷歌学术
  • 夏尔巴·罗密欧
  • 学术期刊数据库
  • 打开 J 门
  • Genamics 期刊搜索
  • 期刊目录
  • 研究圣经
  • 中国知网(CNKI)
  • 乌尔里希的期刊目录
  • 电子期刊图书馆
  • 参考搜索
  • 研究期刊索引目录 (DRJI)
  • 哈姆达大学
  • 亚利桑那州EBSCO
  • OCLC-世界猫
  • 学者指导
  • SWB 在线目录
  • 虚拟生物学图书馆 (vifabio)
  • 普布隆斯
  • 欧洲酒吧
  • ICMJE
分享此页面

抽象的

Identification and Reduction of Matrix Effects Caused by Cremophor EL in Bioanalysis Using Liquid Chromatography/Tandem Mass Spectrometry

Vijaya Bhaskar V, Anil Middha, Sudhir Tiwari and Savithiri Shivakumar

Ion suppression effect of dosing vehicle excipient Cremophor EL (CrEL) on the accuracy of liquid chromatography/ tandem mass spectrometry (LC-MS/MS) measurements was studied. Ion suppression cause significant errors in accuracy of the measured concentrations of test compounds, thereby invalidating the assessment of pharmacokinetic results. Using CrEL as a probe compound, the concentration-time profile of the excipient in plasma from rats dosed both orally and intravenously was determined. The most abundant molecular ions corresponding to PEG oligomers at m/z 828, 872, 916 and 960 with daughter ion at m/z 89 were selected for multiple reaction monitoring (MRM) in electrospray mode of ionization. Plasma concentrations of CrEL ranging from 0.50-1.0 mg/mL in the initial sampling points caused 2-10 fold ion suppression on most of the analytes studied. This can result in false rejection of compounds in a drug discovery screen. In this paper, various sample preparation methods, enhanced chromatographic selectivity, and alternative ionization methods were investigated as means to avoid or minimize ion suppression effects. The elimination of ion suppression effects was achieved by Liquid-Liquid Extraction (LLE) with hexane, TBME in Electrospray Ionisation (ESI) mode as sample preparation method. In contrast to ESI mode that had severe suppression effects from CrEL, atmospheric pressure chemical ionisation (APCI) mode is totally free of suppression effects. The mechanism of ion suppression caused by CrEL in relation to both liquid and gas phase reactions was discussed.

免责声明: 此摘要通过人工智能工具翻译,尚未经过审核或验证。